CranialMask Tracker
Brand(s)
Last updated
Summary
Product
CranialMask Tracker
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| CranialMask Tracker | 6000106896 | 6000-390-000 |
| CranialMask Tracker | 6000106885 | 6000-390-000 |
| CranialMask Tracker | 6000106894 | 6000-390-000 |
Issue
Stryker has discovered an issue where the Cranial Guidance Software may generate the following error message: "Instrument 'Face Mask (6001385000000) is not valid for this application" when the CranialMask Tracker (p/n 6000-390-000) is activated during surgery.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Neurology
Companies
Stryker Leibinger Gmbh & Co. Kg
Botzinger Strasse 41, Freiburg, Germany, 79111
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81286