Health product recall

CT Scan and High Capacity Table

Last updated

Summary

Product
CT Scan and High Capacity Table
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

High Capacity Table

00000222467HM3
00000257352HM5
00000270176HM1
00000256160HM3
00000293344HM8
359928HM9
CL3GT2100005HM
CL3GT1800006HM

5272966
5272966-2
5272966-3

Gt2000x/Vt2000x

00000413151HM2
CL1GT1600075HM
CL1GT1600061HM
00000429212HM4
00000413848HM3
CL1GT1600038HM
000394049HM1
00000429230HM6
00000341402HM6
CL1GT1900057HM
CL1GT2100036HM
CL1GT1800039HM
CL1GT1900060HM
CL1GT1900040HM
CL1GT1900042HM
CL1GT1800049HM
CL1GT2100015HM
HDLVX1900005CN
CL1GT2100091HM
CL1GT1900058HM
CL1GT1800077HM
CL1GT2000019HM
CL1GT1900035HM
CL1GT2100002HM
CL1GT2000062HM
CL1GT1900082HM
00000386645HM6
CL1GT1600029HM
413150HM4
CL1GT2100019HM

5380966

Issue

During a specific servicing activity, if the wrong screws are removed by the service personnel without the appropriate table supports in place, there is the potential for the table to drop.

Recall start date: June 6, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Health products - Medical devices - Radiology
Companies

Ge Hangwei Medical Systems Co. Ltd.

West Area Of Bldg. No.3, Beijing, Beijing, China, 100176

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-64292