Custom-Pak Ophthalmic Surgery Procedure Pack
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Custom-Pak Ophthalmic Surgery Procedure Pack |
More than 10 numbers, contact manufacturer. |
00000-00 TO 99999-99 |
Intrepid Plus FMS Pack, Ultra Infusion Sleeve, Mini-Flared ABS Tip |
More than 10 numbers, contact manufacturer. |
8065752087 |
Centurion Vision System - Active Irrigation FMS Pack, With Intrepid Ultra Sleeves |
More than 10 numbers, contact manufacturer. |
8065752201 |
Turbosonics Tip |
More than 10 numbers, contact manufacturer. |
8065-7900-22 |
Constellation Vision System Fragmentation Pak |
More than 10 numbers, contact manufacturer. |
8065-7509-58 |
Centurion Vision System - Active Irrigation FMS Pack, Infusion Sleeves |
More than 10 numbers, contact manufacturer. |
8065752918 |
Issue
Alcon Canada Inc. will be initiating a voluntary medical device recall (advisory) on specific catalog numbers of Alcon Phaco tips & packs containing those Phaco tips. There is a potential for the Phaco tip wrench to generate plastic particles if the Phaco tip is over-tightened to the handpiece. The purpose of the advisory notice is to provide customer guidance when tightening the Phaco tip to mitigate the risk of generating plastic particles that have potential to enter the eye during surgery.
Recall start date: June 30, 2023
Additional information
Details
Alcon Laboratories Inc.
6201 South Freeway, Fort Worth, Texas, United States, 76134-2099