CyberKnife® Treatment Delivery System
Brand(s)
Last updated
Summary
Product
CyberKnife® Treatment Delivery System
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| CyberKnife® Treatment Delivery System | C0417 | 0660000 |
| CyberKnife® Treatment Delivery System | C0545 | 0660000 |
| CyberKnife® Treatment Delivery System | C0573 | 0660000 |
| CyberKnife® Treatment Delivery System | C0351 | 0660000 |
Issue
During collimator housing drop-off or pickup over the Xchange® table, the system may interlock, stop motion, and prompt service contact. Remove the patient before troubleshooting, do not move the treatment robot using the in-room pendant, and contact accuray service.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Accuray, Inc.
1209 Deming Way, Madison, Wisconsin, United States, 53717
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82239