Discovery MR750w 3.0T Magnetic Resonance Imaging System (2018-06-06)
- Starting date:
- June 6, 2018
- Posting date:
- July 19, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67260
Affected products
Discovery MR750w 3.0T Magnetic Resonance Imaging System
Reason
The following issue is being addressed by GE Healthcare File No. FMI 10908: a small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used.