Health product recall

Discovery NM/CT 670 ES (2018-05-30)

Starting date:
May 30, 2018
Posting date:
July 5, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67166

Affected products

Discovery NM/CT 670 ES

Reason

The following issue is being addressed by GE Healthcare file no. FMI 40875: GE Healthcare has identified that a small number of nuclear systems recently installed with a GE supplied A1 electrical panel may not be properly wired to the partial system ups (uninterruptible power supply).