Drill Generic (2020-08-06)
- Starting date:
- August 6, 2020
- Posting date:
- August 28, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73783
Last updated:
2020-08-28
Affected Products
Drill Generic
Reason
Complaints received from customers that some instruments from additional lots of drill extensions show a reduced or missing clamping functionality.
This instrument is also part of an instruments kit (ref 24980) which may include a single instrument (ref 22520) from affected lots.
Affected products
Drill Generic
Lot or serial number
- 407568
- 407971
- 409386
- 409653
- 411153
- 431849
- 434944
- 435729
Model or catalog number
22520
Companies
- Manufacturer
-
Dentsply Implants Manufacturing GmbH
Rodenbacher Chaussee 4
Hanau
63457
GERMANY