Product
DuraDiagnost
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

DuraDiagnost

Not applicable.

712214

Issue

Philips has identified an issue that may occur with the Philips DXR systems if the collimator is not attached correctly after repairing/replacing. The collimator may fall off during operation which may result in injuries to patients and users.

Recall start date: March 12, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Philips Healthcare (Suzhou) Co., Ltd.
No. 258, Zhong Yuan Road, Suzhou, Jiangsu, China, 215024
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77139

Get notified