DxC 500i Clinical Analyzer and DxC 500 AU Clinical Chemistry Analyzer
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
DxC 500 AU Clinical Chemistry Analyzer |
1.5 or lower |
C63520 |
DxC 500 AU Clinical Chemistry Analyzer |
1.5 or lower |
C63519 |
DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module |
1.5 or lower |
C63521 |
DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module |
1.5 or lower |
C63522 |
DxC 500i Clinical Analyzer, DxC 500i Access 2 Immunoassay Module |
1.5 or lower |
C13252 |
Issue
Beckman Coulter has determined that a modification made by operators to 25 predefined calculated rules, which are included as part of the parameter package, will revert to the default expression following parameter package or software upgrades.