DxC 500i Clinical Analyzer and DxC 500i AU Chemistry Module
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module |
Software 1.5.0 and below |
C63521 |
DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module |
Software 1.5.0 and below |
C63522 |
DxC 500 AU Clinical Chemistry Analyzer |
Software 1.5.0 and below |
C63520 |
DxC 500 AU Clinical Chemistry Analyzer |
Software 1.5.0 and below |
C63519 |
Issue
Beckman Coulter has identified an issue that may affect the reporting of clinical chemistry test results. When a laboratory selects a reporting unit other than the system regional default unit for a clinical chemistry test and multiple tests are ordered under the same sample ID, or multiple replicates of the same test under the same sample ID, the software may incorrectly apply the unit conversion factor more than once. This may result in an erroneous reported value.