Health product recall

DxC 500i Clinical Analyzer and DxC 500i AU Chemistry Module

Last updated

Summary

Product
DxC 500i Clinical Analyzer and DxC 500i AU Chemistry Module
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module

Software 1.5.0 and below

C63521

DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module

Software 1.5.0 and below

C63522

DxC 500 AU Clinical Chemistry Analyzer

Software 1.5.0 and below

C63520

DxC 500 AU Clinical Chemistry Analyzer

Software 1.5.0 and below

C63519

Issue

Beckman Coulter has identified an issue that may affect the reporting of clinical chemistry test results. When a laboratory selects a reporting unit other than the system regional default unit for a clinical chemistry test and multiple tests are ordered under the same sample ID, or multiple replicates of the same test under the same sample ID, the software may incorrectly apply the unit conversion factor more than once. This may result in an erroneous reported value.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Beckman Coulter, Inc.

250 S. Kraemer Blvd., Brea, California, United States, 92821

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82970

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe