Health product recall

Easy Moving Plus B Mobile X-Ray Unit With Digital Detector

Last updated

Summary

Product
Easy Moving Plus B Mobile X-Ray Unit With Digital Detector
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Easy Moving Plus B Mobile X-Ray Unit With Digital Detector

G71467
G77408
G73624
G82557
G79899
G82499

SM-40HF-B-D-VIR

Issue

Notification to inform customers regarding potential safety issue related to the steel cable installed inside the mobile column which supports the weight of telescopic arm with the tube head. The potential risk consists of an interference which could cause that the safety system which blocks the arm of the cable is cut off does not work properly. The interference occurs between the cable and a metal plate of the arm's carriage due to shape of the cable's crimped terminals. Scope of the safety recall is to review the installed base according to the manufacturer's instructions and install the kit immediately if the cable could interfere with the parachute system.

Recall start date: June 5, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Sociedad Espanola De Electromedicina Y Calidad, S.A. (Sedecal)
Poligono Industrial Rio De Janeiro, Algete, Madrid, Spain, 28110
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75707

Get notified

Receive notifications for new and updated recalls and alerts by category.