Ellex Solo-A (2020-04-03)
- Starting date:
- April 3, 2020
- Posting date:
- May 1, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72941
Last updated: 2020-05-01
Affected Products
Ellex Solo-A
Reason
Defect in the joystick wiring. There is the potential that the insulation of the wire can become compromised, potentially causing the exposed conductor to come into contact, with the grounded metal housing. As an outcome when the system is in the "ready" mode, an unintended laser emission may result.
Affected products
Ellex Solo-A
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- LT5106-S
Companies
- Manufacturer
-
ELLEX MEDICAL PTY LTD
3-4 Second Ave.
Mawson Lakes
5095
AUSTRALIA