Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65660

Affected products

Equinox Advantage N2O/O2 Analgesic Gas Mixing and Delivery System

Reason

The manufacturer had been made aware of an incident with one of their Equinox Advantage N2O/O2 mixers where the oxygen supply of the reported mixer (01EQ2000, S/N # EQ 2016-2014) was cut off due to a disengaged piston seat in the equalizer that allowed 100% supply of Nitrous Oxide to be delivered to a patient.

Affected products

Equinox Advantage N2O/O2 Analgesic Gas Mixing and Delivery System

Lot or serial number

EQ2150-2016

Model or catalog number

01EQ2000-DISS