Eterna™ Implantable Pulse Generator
Brand(s)
Last updated
Summary
Product
Eterna™ Implantable Pulse Generator
Issue
Medical devices - Device compatibility
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Eterna™ Implantable Pulse Generator | More than 10 numbers, contact manufacturer. | 32400 |
Issue
Abbott is notifying customers of a potential issue affecting Eterna™ spinal cord stimulation (SCS) implantable pulse generators (IPGs), Model 32400. These IPGs may experience a loss of communication with the clinician programmer (CP) and/or the patient controller (PC). This issue has been traced to Bluetooth® low energy (BLE) circuit component manufacturing.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Neurology
Companies
Abbott Medical
6901 Preston Road, Plano, Texas, United States, 75024
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-78298
Get notified
Receive emails about new and updated recall and safety alerts.