Health product recall

ETMS 70 mm Cooled Coil

Last updated

Summary

Product
ETMS 70 mm Cooled Coil
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Industry

Affected products

Affected products Lot or serial number Model or catalogue number
ETMS 70 mm Cooled Coil CC70-001000-15-P-00007 CC70001
ETMS 70 mm Cooled Coil CC70-001000-15-P-00006 CC70001

Issue

A potential issue manifests as an electrical arc within the coil itself that, when occurs, leads to a relatively loud popping sound,  which can startle the patient or operator.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Rogue Research Inc.
300-6666 Rue Saint-Urbain, Montreal, Quebec, Canada, H2S 3H1
Published by
Health Canada
Audience
Industry
Recall class
Type II
Recall date
Identification number
RA-82196