FreeO2 Automated Oxygen Therapy Device Adult & Pediatric - Version HW 2.3
Brand(s)
Last updated
Summary
Product
FreeO2 Automated Oxygen Therapy Device Adult & Pediatric - Version HW 2.3
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
FreeO2 Automated Oxygen Therapy Device Adult & Pediatric |
Version HW 2.3 |
FO2-110-00 |
Issue
Under specific conditions following a deep battery discharge, error 29515 may occur during startup of a FreeO2 HW 2.3 device using software package 1.4.2 and prevent its use. The update to software package 1.4.3 corrects this issue and restores the expected startup behaviour.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
| Oxynov, Inc. |
| 725, Boulevard Lebourgneuf, Quebec, Quebec, Canada, G2J 0C4 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82749
Get notified