Health product recall

FreeO2 Automated Oxygen Therapy Device Adult & Pediatric - Version HW 2.3

Brand(s)
Last updated

Summary

Product
FreeO2 Automated Oxygen Therapy Device Adult & Pediatric - Version HW 2.3
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

FreeO2 Automated Oxygen Therapy Device Adult & Pediatric

Version HW 2.3

FO2-110-00

Issue

Under specific conditions following a deep battery discharge, error 29515 may occur during startup of a FreeO2 HW 2.3 device using software package 1.4.2 and prevent its use. The update to software package 1.4.3 corrects this issue and restores the expected startup behaviour.
 

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Oxynov, Inc.
725, Boulevard Lebourgneuf, Quebec, Quebec, Canada, G2J 0C4
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82749

Get notified