Health product recall

G7 Acetabular Shell 3 Hole Osseoti, Cementless

Last updated

Summary

Product
G7 Acetabular Shell 3 Hole Osseoti, Cementless
Issue
Medical devices - Labelling and packaging
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

G7 Acetabular Shell 3 Hole Osseoti, Cementless

7314444
7300657
7300679
7304410
7304424
7300682
7300636

110010243
110010245
110010244
110010242

Issue

Zimmer Biomet is conducting a lot specific medical device recall for of various G7 acetabular shells and liner due the outer sterile cavity having the potential to be below the minimum wall thickness. The thin corner wall could potentially crack during transit leading to various issues. Signs of damage to the corner of the blister would be easily recognizable to the end user prior to its usage.

Recall start date: October 4, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies

Biomet Orthopedics

56 East Bell Drive, Warsaw, Indiana, United States, 46581

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72014

Get notified

Receive notifications for new and updated recalls and alerts by category.