Gastrointestinal Videoscope
Brand(s)
Last updated
Summary
Product
Gastrointestinal Videoscope
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Gastrointestinal Videoscope |
More than 10 numbers, contact manufacturer. |
GIF-1100 |
Issue
It was discovered during device performance testing that the CCD imaging sensors were programmed with the incorrect color correction data and therefore, specifications are not met.
Recall start date: September 20, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Companies
Olympus Medical Systems Corp. |
2951 Ishikawa-Cho, Hachioji-Shi, Tokyo-To, Japan, 192-8507 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76246
Get notified