Health product recall

Hamilton-C2 Intensive Care Ventilator

Last updated

Summary

Product
Hamilton-C2 Intensive Care Ventilator
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Hamilton-C2 Intensive Care Ventilator All lots. 160001

Issue

Hamilton Medical AG became aware of a malfunction with HAMILTON-C2 ventilators when used long term with neonatal patients. If an affected HAMILTON-C2 ventilator is used cumulatively for 91 days with neonatal patient without a restart of the device, the device will switch to "ambient state", resulting in an immediate stop of ventilation. During the "ambient state" the device is alarming audibly and visually until it is switched off by the user. Hamilton Medical AG previously communicated this issue through FSCAs with reference CER 110972 (May 2023) and FSCA-2025-01-01 (January 2025). This updated FSN is provided with a new reference number and addresses only HAMILTON-C2 where the neonatal option is installed. Since the initiation of the first field action in May 2023,  Hamilton Medical AG did not receive any new complaint related to the issue for HAMILTON-C2 devices.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Hamilton Medical Ag
Via Crusch 8, Bonaduz, Switzerland, 7402
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Recall date
Identification number
RA-81503