HeartSine® Samaritan® PAD 350P/360P/500P
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
HeartSine® Samaritan® PAD 350p |
More than 10 numbers, contact manufacturer. |
SAM 350P |
HeartSine® Samaritan® PAD 500p |
More than 10 numbers, contact manufacturer. |
SAM 500P |
HeartSine® Samaritan® PAD 360p |
More than 10 numbers, contact manufacturer. |
SAM 360P |
Issue
Stryker has identified a potential issue with the electrical connection between Pad-Pak ™ or Pediatric-Pak™ and the AED. The connection uses spring-loaded electrical contacts, also known as pogo pins. Over time, oxidation of the pogo pin may increase electrical resistance and weaken the internal spring. If the condition progresses, the electrical connection between the Pad-Pak ™ or Pediatric-Pak™ battery and the AED may become intermittent or may be lost.
Additional information
Details
Heartsine Technologies, Ltd.
207 Airport Road West, Belfast, Northern Ireland, United Kingdom, BT3 9ED