Health product recall

HeartSine® Samaritan® PAD 350P/360P/500P

Last updated

Summary

Product
HeartSine® Samaritan® PAD 350P/360P/500P
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected products

Lot or serial number

Model or catalogue number

HeartSine® Samaritan® PAD 350p

More than 10 numbers, contact manufacturer.

SAM 350P

HeartSine® Samaritan® PAD 500p

More than 10 numbers, contact manufacturer.

SAM 500P

HeartSine® Samaritan® PAD 360p

More than 10 numbers, contact manufacturer.

SAM 360P

Issue

Stryker has identified a potential issue with the electrical connection between Pad-Pak ™ or Pediatric-Pak™ and the AED. The connection uses spring-loaded electrical contacts, also known as pogo pins. Over time, oxidation of the pogo pin may increase electrical resistance and weaken the internal spring. If the condition progresses, the electrical connection between the Pad-Pak ™ or Pediatric-Pak™ battery and the AED may become intermittent or may be lost.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Heartsine Technologies, Ltd.

207 Airport Road West, Belfast, Northern Ireland, United Kingdom, BT3 9ED

Published by
Health Canada
Audience
General public
Recall class
Type I
Recall date
Identification number
RA-83012

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe