Health product recall

Hem-o-lok Automatic Ligating Clip Applier (With Clips) - Endoscopic & Weck Auto Endo5 Automatic Hem-o-lok Clip Applier (2019-10-07)

Starting date:
October 7, 2019
Posting date:
November 1, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71479



Last updated: 2019-11-01

Affected Products

  1. Hem-o-lok Automatic Ligating Clip Applier (With Clips) - Endoscopic
  2. Weck Auto Endo5 Automatic Hem-o-lok Clip Applier

Reason

Teleflex is recalling certain lots of surgical automatic clip appliers due to an increase in misloading and/or jamming related complaints that may be due to a discrepancy in the specification of one of the product components. Misloading and/or jamming may lead to clips falling out of the applier, clip breakage, or the jaws becoming locked in a partially closed position (jammed). In these instances, medical or surgical intervention, or procedural delays, could be required.

Affected products

A. Hem-o-lok Automatic Ligating Clip Applier (With Clips) - Endoscopic

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

543965

Companies
Manufacturer

Teleflex Medical

3015 Carrington Mill BLVD

Morrisville

27560

North Carolina

UNITED STATES


B. Weck Auto Endo5 Automatic Hem-o-lok Clip Applier

Lot or serial number

More than 10 numbers, contact manufacturer. 

Model or catalog number

AE05ML

Companies
Manufacturer

Teleflex Medical

3015 Carrington Mill BLVD

Morrisville

27560

North Carolina

UNITED STATES