hillrom?Liko Multirall 200 Q-Link (2021-01-20)
- Starting date:
- January 20, 2021
- Posting date:
- February 5, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-74935
Last updated:
2021-02-05
Affected Products
hillrom?Liko Multirall 200 Q-Link
Reason
Despite reinforcement of the instructions for use for attachment of Q-link 1 strap lock to MultiRall from the ground level with the Extension Arm accessory, serious incidents/adverse events have continued to occur with drop of patient and lift motor. CARSAT a government organization requests changeout to Q-link 2 and usability studies during design development indicate Q-link 2 design was less prone to incorrect attachment due to use error compared to the Q-link 1.
Affected products
hillrom?Liko Multirall 200 Q-Link
Lot or serial number
6501671, 6502767, 6501613
6502436
6503716, 6503551, 6502802
6507752, 6508155, 6509103
6510908, 6510909, 6503531
Model or catalog number
3130001