Health product recall

hillrom?Liko Multirall 200 Q-Link (2021-01-20)

Starting date:
January 20, 2021
Posting date:
February 5, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-74935



Last updated:
2021-02-05

Affected Products 

hillrom?Liko Multirall 200 Q-Link

Reason

Despite reinforcement of the instructions for use for attachment of Q-link 1 strap lock to MultiRall from the ground level with the Extension Arm accessory, serious incidents/adverse events have continued to occur with drop of patient and lift motor. CARSAT a government organization requests changeout to Q-link 2 and usability studies during design development indicate Q-link 2 design was less prone to incorrect attachment due to use error compared to the Q-link 1.

Affected products

hillrom?Liko Multirall 200 Q-Link

Lot or serial number

6501671, 6502767, 6501613

6502436

6503716, 6503551, 6502802

6507752, 6508155, 6509103

6510908, 6510909, 6503531

Model or catalog number

3130001

Companies
Manufacturer

Liko AB

Nedre Vagen 100

Lulea

97592

SWEDEN