Health product recall

Hydrasense Baby Nasal Care Easydose

Brand(s)
Last updated

Summary

Product
Hydrasense Baby Nasal Care Easydose
Issue
Medical devices - Sterility Issue
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products

Lot or serial number

Model or catalog number

Hydrasense Baby Nasal Care Easydose

G239889
G238402

056219998997
056219261282
056500372369

Issue

Manufacturer recalling specific lots due to the possibility of contamination during manufacturing. Although all products passed their release specification testing, recent routine media fill testing showed contamination with ralstonia picketti, leading to the inability to confirm safety of specific product lots. 

As per manufacturer, while no adverse events have been reported at this time, consumers who believe that their product is from an affected lot should stop using it immediately and contact manufacturer for return. 

Recall start date: April 18, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Ear, nose and throat
Companies

Bayer Inc.

2920 Matheson Blvd. East, Mississauga, Ontario, Canada, L4W 5R6

Published by
Health Canada
Audience
General public
Recall class
Type I
Identification number
RA-75468

Get notified

Receive notifications for new and updated recalls and alerts by category.