INNOVANCE® VWF Ac assay
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
INNOVANCE® VWF Ac assay | All lots. | 10873906 |
INNOVANCE® VWF Ac assay | All lots. | 10487040 |
Issue
Siemens Healthineers has confirmed through the investigation of customer complaints that quality control and patient sample results for INNOVANCE® VWF Ac might be falsely elevated, independent of the system and reagent lot used.
An increased number of complaints for the test INNOVANCE® VWF Ac were received, where customers complained of quality control (qc) recovery issues for control plasma p (cpp). Results were found above the permitted range according to the table of assigned values.
There is a potential of elevated INNOVANCE® VWF Ac patient results where qc is within the permitted range.
Date de début du rappel: le 20 juin 2025
Additional information
Details
Siemens Healthcare Diagnostics Products GmbH
Emil-Von-Behring-Str. 76, Marburg, Germany, 35041