Health product recall

INNOVANCE® VWF Ac assay

Last updated

Summary

Product
INNOVANCE® VWF Ac assay
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products Lot or serial number Model or catalog number
INNOVANCE® VWF Ac assay All lots. 10873906
INNOVANCE® VWF Ac assay All lots. 10487040

Issue

Siemens Healthineers has confirmed through the investigation of customer complaints that quality control and patient sample results for INNOVANCE® VWF Ac might be falsely elevated, independent of the system and reagent lot used.

An increased number of complaints for the test INNOVANCE® VWF Ac were received, where customers complained of quality control (qc) recovery issues for control plasma p (cpp). Results were found above the permitted range according to the table of assigned values.

There is a potential of elevated INNOVANCE® VWF Ac patient results where qc is within the permitted range.

Date de début du rappel: le 20 juin 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Haematology
Companies

Siemens Healthcare Diagnostics Products GmbH

Emil-Von-Behring-Str. 76, Marburg, Germany, 35041

Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-77670