InteleShare™
Brand(s)
Last updated
Summary
Product
InteleShare™
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| InteleShare™ | versions 3.24.3 - 3.25.2 | V3.24 |
Issue
The length measurement tool may provide inaccurate values when used in multi-planar reconstruction when rotations are applied. In this scenario, the measurement tool's text may display inconsistent or incorrect length values. Inaccurate measurement values could compromise diagnostic accuracy and potentially lead to patient harm, including misdiagnosis or inappropriate follow-up.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Intelerad Holdings Usa Ltd.
6th Floor, Raleigh, North Carolina, United States, 27609
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81292