Intellijoint KNEE Software Application (2020-02-11)
- Starting date:
- February 11, 2020
- Posting date:
- March 6, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72487
Last updated: 2020-03-10
Affected Products
Intellijoint KNEE Software Application
Reason
The Intellijoint KNEE System may incorrectly report femoral resection depth due to a fault state identified in the underlying algorithm used to calculate resection depth. Irregular condyle topology, lifting the Probe during condyle definition, or capture of an insufficient number of points may affect the resection depth reported.
Affected products
Intellijoint KNEE Software Application
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
000-0547