IntelliSpace Critical Care And Anesthesia
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
IntelliSpace Critical Care And Anesthesia |
Not applicable. |
VERSION J.06 |
Issue
In ICCA releases J.06.01 and J.06.02, calculations triggered by infusion-pump auto-charting may result in incorrect Intake Totals, Output Totals, Net Body Balance, Length of Stay (LOS) Balance, and related fluid-balance values.
This issue applies only to customers that have configured and use Auto Charting for Infusion Data for a Clinical Unit, with one or both of the following settings enabled:
• Automatic periodic charting of infusion rates and volumes.
• Automatic aperiodic charting of infusion rates and bolus volumes when a change occurs.
Customers that do not have either Auto Charting for Infusion Data setting configured and in use are not impacted by this issue.
Additional information
Details
Philips Medizin Systeme Boblingen GmbH
Hewlett-Packard Strasse 2, Boblingen, Germany, 71034