Health product recall

IntelliSpace Critical Care And Anesthesia

Last updated

Summary

Product
IntelliSpace Critical Care And Anesthesia
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

IntelliSpace Critical Care And Anesthesia

Not applicable.

VERSION J.06

Issue

In ICCA releases J.06.01 and J.06.02, calculations triggered by infusion-pump auto-charting may result in incorrect Intake Totals, Output Totals, Net Body Balance, Length of Stay (LOS) Balance, and related fluid-balance values.
This issue applies only to customers that have configured and use Auto Charting for Infusion Data for a Clinical Unit, with one or both of the following settings enabled:
• Automatic periodic charting of infusion rates and volumes.
• Automatic aperiodic charting of infusion rates and bolus volumes when a change occurs.
Customers that do not have either Auto Charting for Infusion Data setting configured and in use are not impacted by this issue.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Philips Medizin Systeme Boblingen GmbH

Hewlett-Packard Strasse 2, Boblingen, Germany, 71034

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82635