Health product recall

Kangaroo Enteral Feeding Spike Sets (2020-09-23)

Starting date:
September 23, 2020
Posting date:
October 23, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-74163



Last updated:
2020-10-22

Affected Products

A. Kangaroo ePump ENPlus Spike with 1000 mL Flush Bag

B. Kangaroo ePump ENPlus Spike Set

C. Kangaroo Joey Pump Feeding Safety Screw Spike Set DEHP-Free PVC

D. Kangaroo Joey ENPlus Spike Set

E. Kangaroo Joey Safety Screw Spike with 1000 mL Flush Bag

F. Kangaroo Connect ENPlus Spike Set, Non-Sterile

G. Kangaroo Connect Bag Set, Non-Sterile

Reason

Cardinal Health had previously notified customers of a medical device product advisory in December 2019 due to the potential risk of leaks with Kangaroo Enteral Feeding Spike Sets. Cardinal has updated this notification to include all unexpired lots manufactured to date. In addition, Cardinal has added two additional product codes of Kangaroo Connect 500 mL and 1000 mL Bag Sets to the product correction notice. Cardinal continues to receive complaints of leaking at the interface of the tube and spike connector, in addition to complaints of leaking due to PVC tubing disconnecting from the cassette in connect bag sets. Cardinal Health Canada is not aware of any injury reports with respect of the use of affected product in the canadian market as of September 22nd 2020.

Affected products

A. Kangaroo ePump ENPlus Spike with 1000 mL Flush Bag

Lot or serial number

All lots.

Model or catalog number

775100

Companies
Manufacturer

Cardinal Health 200, LLC (carrying on the patient recovery business from Covidien LLC)

15 Hampshire Street

Mansfield

02048

Massachusetts

UNITED STATES


B. Kangaroo ePump ENPlus Spike Set

Lot or serial number

All lots.

Model or catalog number

775659

Companies
Manufacturer

Cardinal Health 200, LLC (carrying on the patient recovery business from Covidien LLC)

15 Hampshire Street

Mansfield

02048

Massachusetts

UNITED STATES


C. Kangaroo Joey Pump Feeding Safety Screw Spike Set DEHP-Free PVC

Lot or serial number

All lots.

Model or catalog number

775759

Companies
Manufacturer

Cardinal Health 200, LLC (carrying on the patient recovery business from Covidien LLC)

15 Hampshire Street

Mansfield

02048

Massachusetts

UNITED STATES


D. Kangaroo Joey ENPlus Spike Set

Lot or serial number

All lots.

Model or catalog number

765559

Companies
Manufacturer

Cardinal Health 200, LLC (carrying on the patient recovery business from Covidien LLC)

15 Hampshire Street

Mansfield

02048

Massachusetts

UNITED STATES


E. Kangaroo Joey Safety Screw Spike with 1000 mL Flush Bag

Lot or serial number

All lots.

Model or catalog number

765100

Companies
Manufacturer

Cardinal Health 200, LLC (carrying on the patient recovery business from Covidien LLC)

15 Hampshire Street

Mansfield

02048

Massachusetts

UNITED STATES


F. Kangaroo Connect ENPlus Spike Set, Non-Sterile

Lot or serial number

All lots.

Model or catalog number

77000FD

Companies
Manufacturer

Cardinal Health 200, LLC (carrying on the patient recovery business from Covidien LLC)

15 Hampshire Street

Mansfield

02048

Massachusetts

UNITED STATES


G. Kangaroo Connect Bag Set, Non-Sterile

Lot or serial number

Lots lower than 19201XXXX.

Model or catalog number

77100FD

77500FD

Companies
Manufacturer

Cardinal Health 200, LLC (carrying on the patient recovery business from Covidien LLC)

15 Hampshire Street

Mansfield

02048

Massachusetts

UNITED STATES