Karl Storz Telescopes
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected prodcuts |
Lot or serial number |
Model or catalogue number |
---|---|---|
Telescopes |
All lots. |
27830KA |
Hopkins Telescope |
All lots. |
27093AA |
Nephro/Chold, Berci Telescope |
All lots. |
27295AA |
Issue
The IFUs for the devices identifiers previously included cleaning methods other than manual cleaning and pre-vacuum steam as the approved reprocessing modalities. The IFU was updated to only include these two methods moving forward; therefore, there is a risk that a reprocessing modality that is not currently cleared could be utilized for reprocessing these devices.
Recall start date: July 25, 2024
Additional information
Details
Karl Storz Se & Co. Kg |
Dr.-Karl-Storz Strasse 34, Tuttlingen, Germany, 78532 |