LAERDAL COMPACT SUCTION UNIT 4 (LCSU 4) (2020-10-20)
- Starting date:
- October 20, 2020
- Posting date:
- November 27, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-74413
Last updated:
2020-11-27
Affected Products
LAERDAL COMPACT SUCTION UNIT 4 (LCSU 4)
Reason
The service life of LCSU 4 units manufactured in the period 9 August 2018 to 6 June 2020 may be reduced due to component failures within the internal vacuum air pump. These component failures will result in the LCSU 4 continuing to operate, but not able to provide any suction. As of Sept 25, 2020, Laerdal had received complaints, from less than two percent of all potentially affected LCSU 4s, involving this failure condition.
Prolonged use of the LCSU 4 at vacuum levels above 300 mmHg, either during device testing or during patient use, may cause an air pump component to fail resulting in loss of all suction. Alternate suction devices or methods will be required to continue suctioning.
Affected products
LAERDAL COMPACT SUCTION UNIT 4 (LCSU 4)
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
880051
880052
880061
880062
881151
881152