LIFECODES HLA-B SSO and HLA-B ERES SSO Typing Kit (2018-03-15)
- Starting date:
- March 15, 2018
- Posting date:
- June 22, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67092
Affected products
A. LIFECODES HLA-B SSO Typing Kit
B. LIFECODES HLA-B ERES SSO Typing Kit
Reason
As a result of a complaint investigation, Immucor GTI Diagnostics, Inc. has determined the serological group for the B*27:08 allele was incorrectly assigned as BW4. The correct serological group assignment for B*27:08 is BW6. This assignment was part of the 3.29.0 allele database update.
Affected products
A. LIFECODES HLA-B SSO Typing Kit
Lot or serial number
- 01117A
- 06036A
Model or catalog number
628915
Companies
- Manufacturer
-
Immucor GTI Diagnostics, Inc.,
20925 Crossroads Circle,
Waukesha
53186
UNITED STATES
B. LIFECODES HLA-B ERES SSO Typing Kit
Lot or serial number
- 01117B
- 06036B
Model or catalog number
628917
Companies
- Manufacturer
-
Immucor GTI Diagnostics, Inc.,
20925 Crossroads Circle,
Waukesha
53186
UNITED STATES