Health product recall

MAGNETOM Flow.Elite MR XB10B-HF02

Last updated

Summary

Product
MAGNETOM Flow.Elite
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

MAGNETOM Flow.Elite

SN 234787, software VB10B-2510

11851911

MAGNETOM Flow.Elite

SN 234789, software VB10B-2510

11851911

Issue

Hotfix syngo MR XB10B-HF02 (NX2510-WSMW) is a bug fix update (delta installation) which includes:
• BioMatrix Head/Neck Pro Coil elements detection fixed for the system configurations with Tim[108x24]
and Tim[108x48]
• Improved data logging to improve error case detection
• Improved general system stability during and after start-up
• Improved stability and reliability of the patient table

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Siemens Healthcare GmbH

Henkestr. 127, Erlangen, Germany, 91052

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82938