Med-Rx Extension Sets
Last updated
Summary
Product
Med-Rx Extension Sets
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Med-Rx Extension Sets |
171033 |
10-1074 |
Issue
The issue was identified during use of the extension set, wherein the tubing disconnected at the y-bifurcation junction, specifically at the intended connection interface, creating a product failure.
Recall Start Date: April 17, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Canadian Hospital Specialties Ltd. Doing Business As Chs
2760 Brighton Road, Oakville, Ontario, Canada, L6H 5T4
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77422
Get notified