Health product recall

MicroScan Pos Combo Panels 29, 33, and 34, MicroScan Pos BP Combo Panel Type 20 (2019-03-20)

Starting date:
March 20, 2019
Posting date:
April 12, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69538



Last updated: 2019-04-12

Affected Products

  1. MicroScan Pos Combo Panels 29, 33, and 34
  2. MicroScan Pos BP Combo Panel Type 20

Reason

Beckman Coulter has confirmed customer reports of insufficient growth observed in growth well in overnight conventional gram-positive panel lots with quality control (QC) organism M. Luteus ATCC 49732. The expected result is growth.

The issue was traced to a subassembly used in the manufacturing of the affected panel lots.

Affected products

  1. MicroScan Pos Combo Panels 29, 33, and 34

Lot or serial number

43749

43753

43754

43756

Model or catalog number

B1017-209

B1017-211

B1017-214

Companies
Manufacturer

Beckman Coulter, Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES


B. MicroScan Pos BP Combo Panel Type 20

Lot or serial number

43754

Model or catalog number

B1017-202

Companies
Manufacturer

Beckman Coulter, Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES