Health product recall

MobileDiagnost wDR

Last updated

Summary

Product
MobileDiagnost wDR
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

MobileDiagnost wDR

Not applicable.

9890 010 8952X

Issue

Philips has become aware of a software login issue associated with specific models of Mobilediagnost wDR systems running on Windows 10 operating system preventing use of the system. This issue occurred because the software certificate protecting the Eleva user login expired on May 15, 2023. Therefore, if the customer logs out or reboots the system, they will be prevented from logging in. The issue is in regard to expiry of 6 certificates of Eleva software by 14th May 2023. Eleva systems is unable to function after these dates. Due to the Eleva certificate expiry, the system will not be able to log in and therefore rendering the system non-usable.

Recall start date: May 31, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Philips Medical Systems DMC Gmbh

Roentgenstrasse 24, Hamburg, Germany, 22335

Published by
Health Canada
Audience
Healthcare
Recall class
Type II