MONACO (2021-09-14)
- Starting date:
- September 14, 2021
- Posting date:
- October 1, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-76553
Last updated:
2021-10-01
Affected Products
MONACO
Reason
Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to be frozen on the plans associated with that image set.
Affected products
MONACO
Lot or serial number
- v 5.40.00
- v 5.40.01
- v 5.40.02
- v 5.51.10
Model or catalog number
- MONACO
Companies
- Manufacturer
-
ELEKTA, INC.
1450 BEALE STRET, SUITE 205
St Charles
63303
Missouri
UNITED STATES