Monterey AL AL Implant Inserter
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Monterey AL AL Implant Inserter |
231639 |
48019140 |
Issue
For the lots in scope of this assessment, there has been a trend of reports from the field that found there was the potential for the gold unlock button to separate from the inserter. Investigation determined that this is due to potential tolerance interference within the design that can lead to the pin not being flush, causing weld penetration to not meet specification and the gold unlock button to separate from the inserter.
Recall Start Date: February 27, 2025
Additional information
Details
Stryker Spine
2 Pearl Court, Allendale, New Jersey, United States, 07401
Get notified
Receive notifications for new and updated recalls and alerts by category.