MOSAIQ
Brand(s)
Last updated
Summary
Product
MOSAIQ
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| MOSAIQ | Version 3.2.1.0 to 3.2.3.2 | VERSION 3.2 |
Issue
When using Anatomical Position Monitoring (APM) during a Comprehensive Motion Management (CMM) unity treatment delivery, if the user performs a baseline shift (BLS) plan and an error occurs during the plan transfer, in some instances, the BLS plan may not load into mosaiq resulting in the original beams being delivered without the new BLS applied.
What you should do
Contact the manufacturer if you require additional information.
- Email: application.support.na@elekta.com
- Tel: +1 855 6935358
- https://www.elekta.com
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Elekta Solutions AB
Hagaplan 4, Stockholm, Sweden, 11368
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82038