Health product recall

Noblus Diagnostic Ultrasound Scanner - Main Unit

Last updated

Summary

Product
Noblus Diagnostic Ultrasound Scanner - Main Unit
Issue
Medical devices - Performance issue
What to do

Contact the importer CHRISTIEINNOMED INC at QUALITE.QUALITY@CHRISTIEINNOMED.COM. 

Audience
Healthcare
Industry

Affected products

Affected products

Lot or serial number

Model or catalog number

Noblus Diagnostic Ultrasound Scanner - Main Unit

K1716 0000
AHM250PS24

NOBLUS SYSTEM

Issue

Fujifilm has become aware that some of the AC adapters of Noblus heat up and cause smoke, in the worst case, when excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable). This mechanical stress would cause failure of the solder joint inside of the AC adapter and generates excessive heat.

Recall Start Date: May 9, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Obstetrics and gynaecology
Companies

Fujifilm Healthcare Corporation

2-1 Shintoyofuta, Kashiwa-Shi, Chiba-Ken, Japan, 277-0804

Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-75624

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe