Health product recall

Novum Iq Lvp

Last updated

Summary

Product
Novum Iq Lvp
Issue
Medical devices - Mechanical defect
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected Products

Lot or serial number

Model or catalog number

Novum Iq Lvp

All lots.

40700BAX

Issue

Baxter Corporation has internally identified a scenario on the Novum IQ Large Volume Infusion Pump (lVP) where 1) an unintended second bolus or 2) an unintended second auto-programming order update may be delivered to the patient. Baxter Corporation has identified an additional scenario where a rapid double key press will result in unintended behavior on the Novum IQ LVP. This issue will occur if the user rapidly double presses the 'OK' key (<200 milliseconds apart) when accepting a value exceeding the drug library soft limit.

Recall start date: December 29, 2021.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63792

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe