Olympus MAJ-1443 and MAJ-1444 Valves
Brand(s)
Last updated
Summary
Product
Olympus MAJ-1443 and MAJ-1444 Valves
Issue
Medical devices - Labelling and packaging
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Air/Water Valves | All lots. | MAJ-1444 |
| Reusable Suction Valve | All lots. | MAJ-1443 |
| Endoscope Reprocessor | All lots. | OER-ELITE |
| OER-Pro Endoscope Reprocessor | All lots. | OER-PRO |
Issue
Olympus is informing that the MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OERPro and OER-Elite automated endoscope reprocessors. This means the MAJ-1443 and MAJ-1444 can no longer be high-level disinfected using the OER-Pro or the OER-Elite and must instead be reprocessed by methods described in updated IFU.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Health products - Medical devices - General hospital and personal use
Companies
Olympus Medical Systems Corp.
2951 Ishikawa-Cho, Hachioji-Shi, Tokyo-To, Japan, 192-8507
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81697