Health product recall

Optilite Low Level Albumin Kit (2019-12-19)

Starting date:
December 19, 2019
Posting date:
January 10, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72081



Last updated: 2020-01-10

Affected Products

Optilite Low Level Albumin Kit

Reason

An internal complaint was raised by a TBS member of staff regarding an error in the barcodes on the product Quality Control certificate provided with the affected kit lot (complaint raised 29/NOV/19). This error presents differently depending on the software version the Optilite instrument is running. The Optilite software version 7 barcodes cannot be used, scanning these will give an error message which cannot be overridden. The Optilite software version 6.0.1 barcodes may lead to an intermittent error message, however this message does not affect the performance of the kit. Users running Optilite version 7 software may scan the software version 6.0.1 barcodes in order to use the affected kit.

Affected products

Optilite Low Level Albumin Kit

Lot or serial number

424266

Model or catalog number

NK032.L.OPT

Companies
Manufacturer

The Binding Site Group Limited

8 Calthorpe Road

Birmingham

B15 1QT