Ortho Vision Analyzer
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Ortho Vision Analyzer |
More than 10 numbers, contact manufacturer. |
6904577 |
Ortho Vision Max Analyzer |
Not applicable. |
6153434 |
Ortho Vision Max Analyzer |
More than 10 numbers, contact manufacturer. |
6904576 |
Issue
The loading station door unlocks for 20 seconds upon request to load samples or diluents. If the door is not opened within this period, it relocks without triggering a re-inventory since no user access occurred. If the door is opened immediately after the allotted 20-second countdown ends but within the door's auto-relock timeframe, the user will be allowed to load samples, however, re-inventory will not be triggered as the software does not detect that the loading station door was opened. There are no error codes generated for this issue. An inventory event will take place the next time you properly access the loading station.
Recall Start Date: April 25, 2025
Additional information
Details
Ortho-Clinical Diagnostics Inc.
1001 Us Hwy 202, Raritan, New Jersey, United States, 08869
Get notified
Receive notifications for new and updated recalls and alerts by category.