Health product recall

Osstell® Smartpeg Type 101-104

Brand(s)
Last updated

Summary

Product
Osstell® Smartpeg Type 101-104
Issue
Medical devices - Performance
What to do

Contact the manufacturer Osstell has a new IFU under implementation and has initiated a design change. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Osstell® Smartpeg Type 101-104 Not applicable. 100766
Osstell® Smartpeg Type 101-104 Not applicable. 100767
Osstell® Smartpeg Type 101-104 Not applicable. 100769
Osstell® Smartpeg Type 101-104 Not applicable. 100768

Issue

If not correctly used with a reference measurement on implant level. The measurement on abutment can indicate a false high ISQ value. This can influence the clinician to take the wrong decision and load the implant too early, in the extension potentially resulting in implant failure for the patient.

Recall start date: May 23, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Dental
Companies

Osstell AB

Stampgatan 14, Goteborg, Sweden, 41101

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77736

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe