Oxygo Portable Oxygen Concentrator (2019-08-16)
- Starting date:
- August 16, 2019
- Posting date:
- August 30, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70853
Last updated: 2019-09-05
Affected Products
Oxygo Portable Oxygen Concentrator
Reason
A potential abrupt shut down of the oxygen concentrator may occur due to an over-draw when a fully charged 8-cell battery is connected to the oxygen concentrator as its only power source. The oxygen concentrator may become unresponsive for up to one minute, after which it will behave as though a battery was just connected with standard flash screen on display.
Affected products
Oxygo Portable Oxygen Concentrator
Lot or serial number
SN's: 19223909-19256614
Model or catalog number
1400-1000
Companies
- Manufacturer
-
Applied Home Healthcare Equipment
28825 Ranney Parkway
Westlake
44145
Ohio
UNITED STATES