Philips Breast Coils
Last updated
Summary
Product
Philips Breast Coils
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
LBS Breast Array Coil |
Not applicable. |
4535-302-8072X |
dS Breast |
Not applicable. |
4598-017-29881 |
Issue
If the cross-section of the prone patient, breast coil, and patient table exceeds the internal diameter of the magnet bore, the patient may be compressed between the breast coil and top of the magnet bore potentially resulting in harm to the patient.
Recall start date: November 7, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Philips Medical Systems Nederland B.V.
Veenpluis 6, Best, Netherlands, 5684 PC
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76881
Get notified