Health product recall

Philips Bridge Prep Kit

Last updated

Summary

Product
Philips Bridge Prep Kit
Issue
Medical devices - Labelling and packaging
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Bridge Prep Kit

More than 10 numbers, contact manufacturer.

591-001

Issue

Potential device interaction issue associated with the use of the bridge prep kit with the bridge occlusion balloon catheter. Specifically, in some cases, the catheters may experience resistance when being advanced over the 0.035" InQwire® Amplatz Guide Wire supplied in the bridge prep kit.  This impacts unexpired lots of the bridge prep kit.  Note: the bridge occlusion balloon is within specifications and does not present any issues when used with a conforming 0.035" guidewire.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Philips Image Guided Therapy Corporation

9965 Federal Drive, Colorado Springs, Colorado, United States, 80921

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82981