Health product recall

Power X2 Powered Ambulance Cot

Brand(s)
Last updated

Summary

Product
Power X2 Powered Ambulance Cot
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Power X2 Powered Ambulance Cot S/N 25G037093 001-5816
Power X2 Powered Ambulance Cot S/N 25G037597 001-5816
Power X2 Powered Ambulance Cot S/N 25G037598 001-5816

Issue

Ferno has identified an issue wherein the fastening post hardware could loosen allowing the fastening post to detach from the cot. The fastening post is intended to secure the ambulance cot in the ambulance fastener during routine patient transport. An unsecured ambulance cot could cause injury to patient and or medic.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Physical medicine
Companies
Ferno-Washington
70 Weil Way, Wilmington, Ohio, United States, 45177
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Recall date
Identification number
RA-78097

Get notified

Receive notifications for new and updated recalls and alerts by category.