Pressure Slave Cable (2018-05-15)
- Starting date:
- May 15, 2018
- Posting date:
- June 6, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-66978
Affected products
Pressure Slave Cable
Reason
Pressure Slave Cable part no. 650404-004, model number 70anm, contains a capacitor part no. 851525026 that is not RoHS compliant. The European Union Directive 2011/65/EU on the restriction of hazardous substances (RoHS II) restricts the use of specific hazardous materials found in electrical and electronic products. The capacitor inside of slave cable contains 40% lead.
Affected products
Pressure Slave Cable
Lot or serial number
S1506036
Model or catalog number
70ANM5
Companies
- Manufacturer
-
Edwards Lifesciences LLC
1 Edwards Way
Irvine
92614
UNITED STATES