PRO-NOX Nitrous Oxide Delivery System (2018-05-09)
- Starting date:
- May 9, 2018
- Posting date:
- August 17, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67444
Affected products
PRO-NOX Nitrous Oxide Delivery System
Reason
It was found that a couple of PRO-NOX O2/N2 mixers were reported to have delivered levels of nitrous oxide higher than specifications during routine testing. No patients were involved. This recall action taken to perform upgrades to the units in the market.
Affected products
PRO-NOX Nitrous Oxide Delivery System
Lot or serial number
Contact manufacturer
Model or catalog number
CMI-0100-PNX